Media Autoclave 2020
A Tender Notice
by MEDICINES HEALTHCARE PRODUCTS REGULATORY AGENCY
- Source
- Find a Tender
- Type
- Contract (Supply)
- Duration
- 2 month
- Value
- 50K
- Sector
- INDUSTRIAL
- Published
- 07 Jan 2021
- Delivery
- To 08 Apr 2021 (est.)
- Deadline
- 07 Feb 2021 23:00
Concepts
Location
National Institute for Biological Standards and Control Blanche Lane, South Mimms, Potters Bar EN6 3QG
2 buyers
Description
A replacement autoclave is required for sterilisation of media and equipment, the current machine (LTE Touchclave R) is old and not fit for purpose due to the time taken to complete each cycle and the capacity / shape of the chamber. Current Cycles required are as shown in the table below (the machine must have capacity for additional cycles to be added)
Total Quantity or Scope
The new autoclave is required to have: • Chamber capacity of at least 200 litres • Overall dimensions, depth no larger than 1200mm (approx. dimensions to be ~1800mm high x 800 mm wide x 1200mm deep) • A removable shelf • Armoured ballast (product) probe • Cooling jacket or similar means of cooling items at the end of the cycle • Cooling cycle must take no longer than 60 minutes for any load • A paperless chart recording with the function of being connected via network cable (via CAT 7 data point) to Eurotherm Software • Linked to a paper printer as a back up to the paperless system • Programming must be open code (including autoclave cycles to allow them to be adjusted by competent staff) • Easily accessible to allow servicing and changing of components to comply with all relevant standards • All probes must be calibrated to UKAS standards.
CPV Codes
- 38000000 - Laboratory, optical and precision equipments (excl. glasses)
Indicators
- Bids should cover the whole contract.
- Renewals are not available.
- Award on basis of price and quality.
Reference